ECP Clinical Effect Comparison Between Positive/Negative Pressure Type vs Positive Pressure Type 2015-1-7


 

ECP Clinical Effect Comparison Between Positive/Negative Pressure Type vs Positive Pressure Type 
 

ECP Device Effect Comparison on Hemodynamic Changes Between Positive/Negative Pressure Type vs Positive Pressure Type
 

Cardiac Function Department of Second Clinical Medical School Affiliated Xian Medical University 
 

Sun Jichuan, Tian Qinghai, Wang Shichen


 In order to observe the effect of Positive/Negative ECP device (model WFB-V) applied in Hemodynamic, 9 cases (test group) who received ECP treatment were studied through catheter method on Hemodynamic monitoring, thereof 5 cases are male, and 4 cases are female, age 50±11.0. They were judged 3 cases sinus bradycardia, 3 cases sick sinus, 2 cases coronary heart disease, 1 case palpitations. There were 14 cases in control group (applied positive type ECP device), thereof 7 cases are male, and 7 cases are female, age 52.4±13.2, they were judged 3 cases sinus bradycardia, 5 cases sick sinus, 2 cases coronary heart disease, 4 case others. Both 2 groups applied aortic root catheter pressure and dish - C z right cardiac catheterization pressure. ECP treated 1 hour, monitoring the parameters changes before, during and after ECP treatment. Statistical processed and comparison. The following tables are the parameters comparison between two types ECP devices treatment.
 
   
Table 1  The change of the aortic systolic pressure    unit: mmHg

group

Cases

ECP 60’ vs before ECP

after ECP vs before ECP

Positive Type

14

decreased 5.87

decreased 7.72

Positive/Negative type

9

decreased 25.60

decreased 9.77

Note1mmHg=01333kPa
      Table 2  The level of aortic regurgitation pressure    unit: mmHg

group

cases

during ECP treatment value

Positive Type

14

131.30~132.69

Positive/Negative type

9

117.84~125.78

           
      Table 3  The aortic mean pressure       unit: mmHg

group

cases

ECP 60’ vs before ECP

after ECP vs before ECP

Positive Type

14

increased13.15

increased 0.38

Positive/Negative type

9

increased 476

decreased 8.95

               
       Table 4  The change of the aortic diastolic pressure    unit: mmHg 

group

cases

ECP 60’ vs before ECP

after ECP vs before ECP

Positive Type

14

increased 7.61

increased 0.55

Positive/Negative type

9

decreased 1.22

decreased 1.55

         
     Table 5  The change of the pulmonary artery systolic pressure    unit: mmHg

group

cases

ECP 60’ vs before ECP

after ECP vs before ECP

Positive Type

14

increased 0.38

decreased 3.34

Positive/Negative type

9

increased 0.38

decreased 3.44

Table 6   The change of the pulmonary artery diastolic pressure  unit: mmHg

group

cases

ECP 60’ vs before ECP

after ECP vs before ECP

Positive Type

14

decreased 0.22

decreased 1.56

Positive/Negative type

9

no change

decreased 3.53

               
     Table 7   The change of pulmonary capillary pressure    unit: mmHg

group

cases

ECP 60’ vs before ECP

after ECP vs before ECP

Positive Type

14

decreased 0.24

decreased 1.86

Positive/Negative type

9

increased 0.18

decreased 2.88

               
     Table 8   Central venous pressure changes    unit: mmHg

group

cases

ECP 60’ vs before ECP

after ECP vs before ECP

Positive Type

14

increased 0.66

decreased 0.55

Positive/Negative type

9

increased 2.29

decreased 1.1

               
   
     Through the table 1-8, you can see the effect from the 9 cases applied Huanan brand positive/negative ECP device (model WFB-V): 1The decrease of the aortic systolic pressure is more obvious in positive/negative type than that in positive type(mean decrease value reached 25.6nanltg). and decrease after stop ECP is still more than in positive type(9.77mmHg);  2The level of aortic regurgitation pressure, its increment is a little lower than in positive type, but its aortic mean pressure is significant decrease after stop ECP than before ECP. 3 the unique and outstanding feature is the positive/negative type ECP all can decreased aortic diastolic pressure during ECP or after ECP, but the positive type increased on the contrary; 4two type ECP are similar in pulmonary artery systolic pressure, pulmonary capillary pressure and central venous pressure, but the positive/negative type decreased obviously in pulmonary artery systolic pressure, pulmonary capillary pressure and central venous pressure after stop ECP treatment. All above 4 points indicate that positive/negative type ECP device can decrease heart load during treatment and improved the heart function significantly.
This paper is one of the documents of Mechanism of external counterpulsation device treatment series research of science and technology achievement appraisal ,  organizing appraisal unit belonged to the Food and Drug Administration of Guangdong province.
 
 